Objective breast symmetry analysis with the breast analyzing tool (BAT) : improved tool for clinical trials
Verfasser / VerfasserinKrois, Wilfried ; Romar, Alexander Ken ; Wild, Thomas ; Dubsky, Peter ; Exner, Ruth ; Panhofer, Peter ; Jakesz, Raimund ; Fitzal, Florian
Erschienen in
Breast Cancer Research and Treatment, 2017, Jg. 164, H. 2, S. 421-427
ErschienenSpringer, 2017
SpracheEnglisch
DokumenttypAufsatz in einer Zeitschrift
Schlagwörter (EN)Breast symmetry / Objective breast analysis / Breast cosmetics / Breast analyzing tool
URNurn:nbn:at:at-ubmuw:3-3526 Persistent Identifier (URN)
DOI10.1007/s10549-017-4255-z 
Zugriffsbeschränkung
 Das Werk ist frei verfügbar
Dateien
Objective breast symmetry analysis with the breast analyzing tool (BAT) [0.89 mb]
Links
Nachweis
Klassifikation
Zusammenfassung (Englisch)

Purpose

Objective cosmetic analysis is important to evaluate the cosmetic outcome after breast surgery or breast radiotherapy. For this purpose, we aimed to improve our recently developed objective scoring software, the Breast Analyzing Tool (BAT®).

Methods

A questionnaire about important factors for breast symmetry was handed out to ten experts (surgeons) and eight non-experts (students). Using these factors, the first-generation BAT® software formula has been modified and the breast symmetry index (BSI) from 129 women after breast surgery has been calculated by the first author with this new BAT® formula. The resulting BSI values of these 129 breast cancer patients were then correlated with subjective symmetry scores from the 18 observers using the Harris scale. The BSI of ten images was also calculated from five observers different from the first author to calculate inter-rater reliability. In a second phase, the new BAT® formula was validated and correlated with subjective scores of additional 50 women after breast surgery.

Results

The inter-rater reliability analysis of the objective evaluation by the BAT® from five individuals showed an ICC of 0.992 with almost no difference between different observers. All subjective scores of 50 patients correlated with the modified BSI score with a high Pearson correlation coefficient of 0.909 (p < .001) which was better compared to the old software (r = 0.769; p < .001).

Conclusions

The modified BAT® software improves the correlation between subjective and objective BSI values, and may be a new standard for trials evaluating breast symmetry.